logo_small.gif (4090 Byte)

Home
About
Products
Software
Literature
Prices
Conferences
Contact

from the content:

Photon Counting Performance of Selected Detectors

R3809U MCP-PMT

The TCSPC system response for a Hamamatsu R3809U-50 MCP [27] is shown in fig. 20. The MCP was illuminated with a femtosecond Ti:Sa laser, the response was measured with an SPC-630 TCSPC module. A HFAC-26-01 preamplifier was used in front of the SPC-630 CFD input. At an operating voltage of -3 kV the FWHM (full width at half maximum) of the response is 28 ps.

Linear scale

Logarithmic scale

Fig. 20 R3809U, TCSPC instrument response. Operating voltage-3kV, preamplifier gain 20dB, discriminator threshold - 80mV

The response has a shoulder of some 400 ps duration and about 1% of the peak amplitude. This shoulder seems to be a general property of all MCPs and appears in all of these devices. The width of the response can be reduced to 25 ps by increasing the operating voltage to the maximum permitted value of -3.4 kV. However, for most applications this is not recommended for the following reason:

Fig. 21 shows that count rates in excess of 2 MHz can be reached.

Fig. 22 shows the histogram of the time intervals between the recorded photons. The count rate was about 10,000 photons per second, the data were obtained with an SPC-830 in the ‘FIFO’ mode. Interestingly, the R3809U is free of afterpulsing.

As all MCP-PMTs, the R3809U allows only a very small maximum output current. This sets a limit to the maximum count rate that can be obtained from the device. The maximum count rate depends on the MCP gain, i.e. of the supply voltage. The count rate for the maximum output current of 100 nA as a function of the supply voltage is shown in fig. 21. To keep the counting efficiency constant the CFD threshold was adjusted to get a constant count rate at a reference intensity that gave 20,000 counts per second. The R3809U tubes have a relatively good SER pulse height distribution which seems to be independent of the cathode type - possibly a result of the independent manufacturing of the channel plate and the cathode. Therefore a good counting efficiency can be achieved.

Due to the short TCSPC response and the absence of afterpulses the R3809U is an ideal detector for TCSPC fluorescence lifetime measurements, for TCSPC lifetime imaging, and for combined lifetime / FCS or other correlation experiments. Recently Hamamatsu announced the R3809U MCP with GaAs, GaAsP, and infrared cathodes for up to 1700 nm. Although these MCPs are not as fast as the versions with conventional cathodes they might be the ultimate detectors for combined FCS / lifetime experiments.

The flipside is that MCPs are expensive and can easily be damaged by overload. Therefore the R3809U should be operated with a preamplifier that monitors the output current. If overload conditions are to be expected, i.e. by the halogen or mercury lamp of a scanning microscope, electronically driven shutters should be used and high voltage shutdown should be accomplished to protect the detector.

H7422

The H7422 incorporates a GaAs or GaAsP cathode PMT, a thermoelectric cooler, and the high voltage power supply [28]. Hamamatsu delivers a small OEM power supply to drive the cooler. However, we could not use this power supply because it generated so much noise that photon counting with the H7422 was not possible. Furthermore, we found that the H7422 shuts down if the gain control voltage is changed faster that about 0.1V / s. Apparently fast changes activate an internal overload shutdown. Unfortunately the device can only be re-animated by cycling the +12 V power supply.

Therefore we use the Becker & Hickl DCC-100 detector controller. It drives the cooler and supplies the +12 V and a software-controlled gain control voltage to the H7422. Furthermore, the DCC in conjunction with a HFAC-26-1 preamplifier can be connected to shut down the gain of the H7422 on overload. If the H7422 shuts down internally for any reason, cycling the +12 V is only a mouse click into the DCC-100 operating panel.

Linear scale

Logarithmic scale


Fig. 23: H7422-40, TCSPC Instrument response function. Gain control voltage 0.9V (maximum gain), preamplifier 20dB, discriminator threshold -200mV, -300mV, -400mV and -500mV, linear scale

The FWHM of the system response is about 300 ps. There is a weak secondary peak about 2.5 ns after the main peak, and a peak prior to the main peak can appear at low discriminator thresholds. The width of the response does not depend appreciably of the discriminator threshold. This is an indication that the response is limited by the intrinsic speed of the semiconductor photocathode.

Fig. 24: H7422-40, Histogram of times between photons. Gain control voltage 0.9V (red)  0.63V (blue). Afterpulse probability increases with gain.

Fig. 25: H7422-40, TCSPC Instrument response function. Gain control voltage 0.63 V, preamplifier 20dB, discriminator threshold -30 mV, -50 mV, -70 mV

The afterpulsing probability of the H7422-40 can be seen from the histogram of the time intervals of the photon (fig. 24). For maximum gain the afterpulse probability in the first 1.5 µs is very high (fig. 24, red curve, control voltage 0.9V). If the gain is reduced the afterpulse probability decreases considerably (fig. 24, blue curve, 0.63V). The timing resolution does not decrease appreciably at the reduced gain, fig. 25.

The H7422 is a good detector for TCSPC applications when sensitivity has a higher priority than time resolution. A typical application is TCSPC imaging with laser scanning microscopes [18,29]. The high quantum efficiency helps to reduce photobleaching which is the biggest enemy of lifetime imaging in scanning microscopes.

The H7422 can also be used to investigate diffusion processes in cells or conformational changes of dye / protein complexes by combined FCS / lifetime spectroscopy. Although the accuracy in the time range below 1.5 µs is impaired by afterpulsing, processes at longer time scales can be efficiently recorded.

Another application of the H7422 is optical tomography with pulses NIR lasers. Because the measurements are run in-vivo it is essential to acquire a large number of photons in a short measurement time. Particularly in the wavelength range above 800 nm the efficiency of H7422-50 and -60 yields a considerable improvement compared to PMTs with conventional cathodes.

H7421

The Hamamatsu H7421 is similar to the H7422 in that it contains a GaAs or GaAsP cathode PMT, a thermoelectric cooler, and the high voltage power supply. However, the output of the PMT is connected to a discriminator that delivers TTL pulses. The output of the PMT is not directly available, and the PMT gain and the discriminator threshold cannot be changed. The module is therefore easy to use. However, because the discriminator is not of the constant fraction type, the TCSPC timing performance is by far not as good as for the H7422, see figure 26.

Linear scale

Logarithmic scale


Fig. 26:
H7422-50, TCSPC response function for a count rate of 30 kHz (blue) and 600 kHz (red), logarithmic scale scale

The FWHM is only 600 ps. Moreover, it increases for count rates above some 100 kHz. Interestingly no such count rate dependence was found for the H7422. Obviously the H7422 is a better solution if high time resolution and high peak count rate is an issue.

H5783 and H5773 Photosensor Modules, PMH-100

The H5783 and H5773 photosensor modules contain a small (TO9 size) PMT and the high voltage power supply [30]. They come in different cathode and window versions. A ‘P’ version selected for good pulse height distribution is available for the bialkali and multialkali tubes. The typical TCSPC response of a H5773P-0 is shown in fig. 27. The device was tested with a 650 nm diode laser of 80 ps pulse width. A HFAC-26-10 preamplifier was used, and the response was recorded with an SPC-730 TCSPC module.

The response function has a pre-peak about 1 ns before the main peak and an secondary peak 2 ns after. The pre-peak is caused by low amplitude pulses, probably from photoemission at the first dynode. It can be suppressed by properly adjusting the discriminator threshold. The secondary peak is independent of the discriminator threshold.

Linear scale

Logarithmic scale


Fig. 27:
H5773P-0, TCSPC instrument response. Maximum gain, preamplifier gain 20dB, discriminator threshold -100mV, -300mV and -500mV, logarithmic scale

The Becker & Hickl PMH-100 module contains a H5773P module, a 20 dB preamplifier, and an overload indicator. The response is the same as for the H5773P and a HFAC-26 amplifier. However, because the PMT and the preamplifier are in the same housing, the PMH-100 has a superior noise immunity. This results in an exceptionally low differential nonlinearity in TCSPC measurements.


Fig. 28:
H7422-40, Histogram of times between photons for H5773P-0 (blue) and H5773P-1 (red). The afterpulse probability is higher for the -1 version.
Fig.29:  Dark count rate for different H5773P-1 modules as a function of ambient temperature.

A histogram of the times between the photon pulses for the H5773 is shown in fig. 28. The devices show relatively strong afterpulsing, particularly the multialkali (-1) tubes. Taking into regards the small cathode area of the devices the dark count rates are relatively high. Selected devices with lower dark count rate are available.

The H5783, the H5773 and particularly the PMH-100 are easy to use, rugged and fast detectors that can be used for TCSPC, multiscalers and gated photon counting as well. In multiscaler applications the detectors reach peak count rates of more that 150 MHz for a few 100 ns. The detectors are not suitable for FCS or similar correlation experiments on the time scale below 1 us.

R7400 and R5600 TO-8 PMTs

The R7400 and the older R5600 are bare tubes similar to that used in the H5783 and H5773. There is actually no reason to use the bare tubes instead of the complete photosensor module. However, for the bare tube the voltage divider can be optimised for smaller TTS or improved linearity at high count rate. The TTS width decreases with the square root of the voltage between the cathode and the first dynode. It is unknown how far the voltage can be increased without damage. A test tube worked stable at 1 kV overall voltage with a three-fold increase of the cathode-dynode voltage. The decrease of the response width is shown in fig. 30.


Fig. 30: R5600P-1, -1kV supply voltage: TCSPC response for different voltage between cathode and first dynode. photons. Blue, green and red: 1, 2 and 3 times nominal voltage

Fig. 31: H5773P-1, -1kV : Histogram of times between photons

The afterpulse probability is the same as for the H5783 and H5773 photosensor modules (fig. 31). It is questionable whether the benefit of a slightly shorter response compensates for the inconvenience of building a voltage divider and using a high voltage power supply. However, if a large number of tubes has to be used, i.e. in an optical tomography setup, using the R5600 or R7400 can be reasonable.

R5900-L16 Multichannel PMT and PML-16 Multichannel Detector Head

The Hamamatsu R5900-L16 is a multi-anode PMT with 16 channels in a linear arrangement. In conjunction with a polychromator the detector can be used for multi-wavelength detection. If the R5900-L16 is used with steady-state and gated photon counters or with multiscalers 16 parallel recording channels, e.g. two parallel Becker & Hickl PMM-328 modules are required. For TCSPC application the multi-detector technique described in [9] and [12-15] can be used. TCSPC multi-detector operation is achieved by combining the photon pulses of all detector channels into one common timing pulse line and generating a ‘channel’ signal which indicates in which of the PMT channels a photon was detected. The Becker & Hickl PML-16 detector head [13] contains the R5900-L16 tube and all the required electronics.

The R5900-L16 has also been used with a separate routing device [12,31]. However, in a setup like this noise pick-up from the environment and noise from matching resistors and preamplifiers adds up so that the timing performance is sub-optimal.

The TCSPC response of two selected channels of the PML-16 detector head is shown in fig. 32. The response of a single channel of different R5900-L16 is between 150 ps and 220 ps FWHM.


Fig. 32: System response of two selected channels of the PML-16 detector head

The response is slightly different for the individual channels. Fig. 33 shows the response for the 16 channels as sequence of curves and as a colour-intensity plot. There is a systematic wobble in the delay of response with the channel number. That means, for the analysis of fluorescence lifetime measurements the instrument response function (IRF) must be measured for all channels, and each channel must be de-convoluted with its individual IRF.


Fig. 33: System response of the PML-16 / R5900-L16 channels. Left curve plot, right colour-intensity plot

The data sheet of the R5900-L16 gives a channel crosstalk of only 3%. There is certainly no reason to doubt about this value. However, in real setup it is almost impossible to reach such a small crosstalk. If crosstalk is an issue the solution is to use only each second channel of the R5900-L16 [31]. If the PML-16 is used with only 8 channels, the data of the unused channels should simply remain unused. If the R5900-L16 is used outside the PML-16 the unused anodes should be terminated into ground with 50 W .


Fig. 34: R5900-L16, histogram of times between photons. No afterpulsing was found.

A histogram of the times between the photon pulses is shown in fig. 34. No afterpulsing was found in the R5900-L16. It appears unlikely that the absence of afterpulses was a special feature of the tested device. The result is surprising because afterpulsing is detectable in all PMTs of conventional design. It seems that the ‘metal channel’ design of the R5900 is really different from any conventional dynode structure. That means, the R5900-L16 and the PML-16 detector head are exceptionally suitable for combined multi-wavelength fluorescence lifetime and FCS experiments. The absence of afterpulses can be a benefit also in high repetition rate TCSPC measurements in that there is no signal-dependent background. A R5900-L16 with a GaAs or GaAsP cathode - although not announced yet - would be a great detector.

Side Window PMTs

Side window PMTs are rugged, inexpensive, and often have somewhat higher cathode efficiency than front window PMTs. The broad TTS and the long SER pulses make them less useful for TCSPC application or for multiscaling or gated photon counting with high peak count rates. However, side-window PMTs are used in many fluorescence spectrometers, in femtosecond correlators and in laser scanning microscopes. If an instrument like these has to be upgraded with a photon counting device it can be difficult to replace the detector. Therefore, some typical results for side window PMTs are given below.

The width and the shape of the TCSPC system response depend on the size and the location of the illuminated spot on the photocathode. The response for the R931 - a traditional 28 mm diameter PMT - for a spot diameter of 3 mm is shown in Fig. 35.

Linear scale

Logarithmic scale


Fig. 35: R931, TCSPC system response for different spots on the photocathode. Spot diameter 3mm

By carefully selecting the spot on the photocathode an acceptable response can be achieved [31,32]. A TCSPC response width down to 112 ps FWHM has been reported [32]. This short value was obtained by using single electron pulses in an extremely narrow amplitude interval and illuminating a small spot near the edge of the cathode.

The afterpulse probability for an R931 is shown in Fig. 36. The afterpulse probability depends on the operating voltage, and the afterpulses occur within a time interval of about 3 µs. The high afterpulse probability does not only exclude correlation measurements on the time scale below 3 µs, it can also result in a considerable signal-dependent background in high repetition rate TCSPC applications


Fig. 36: R931,
histogram of times between photons. Red - 900V, blue -1000V. The afterpulse probability increases with voltage.

Surprisingly, modern 13 mm diameter side window tubes are not faster than the traditional 28 mm tubes. The TCSPC response for a Hamamatsu R6350 is shown in fig. 37.

13 mm tubes are often used in the scanning heads of laser scanning microscopes. It is difficult, if not impossible to replace the side-window PMTs with faster detectors in these instruments. Therefore it is often unavoidable to use the 13 mm side-on tube for TCSPC lifetime imaging. Depending on the size and the location of the illuminated spot an FWHM of 300 to 600 ps can be expected. Although this is sufficient to determine the lifetimes of typical high quantum yield chromophores, accurate FRET and fluorescence quenching experiments require a higher time resolution.

Linear scale

Logarithmic scale

Fig. 37: R6350, TCSPC system response for illumination of full cathode area

CP 944 Channel Photomultiplier

The channel photomultipliers of Perkin Elmer offer high gain and low dark count rates at a reasonable cost. Unfortunately the devices have an extremely broad TTS. The TCSPC system response to a 650nm diode laser is shown in fig. 38. The FWHM of the response is of the order of 1.4 to 1.9 ns which is insufficient for typical TCSPC applications.

Linear scale

Logarithmic scale


Fig. 38: CP 944 channel photomultiplier, TCSPC response. 650 nm, count rate 1.5.105, high voltage -2.8 kV (red) and -2.9 kV (blue). Full cathode illuminated

However, the Perkin Elmer channel PMTs have high gain, a low dark count rate and a surprisingly narrow pulse height distribution. This makes them exceptionally useful for low intensity steady state photon counting or multichannel scaling.

SPCM-AQR Single Photon Avalanche Photodiode Module

The Perkin Elmer SPCM-AQR single photon avalanche photodiode modules are well-known for their high quantum efficiency in the near-infrared. Unfortunately the modules have a very poor timing performance. The TCSPC response for a SPCM-AQR-12 (dark count class <250 cps) is shown in fig. 39.


Fig. 39: SPCM-AQR-12, TCSPC response. Left: 405nm, red 50 kHz, blue 500 kHz count rate. Right: 650 nm, red 50 kHz, blue 500 kHz count rate

The response was measured with a 405 nm BDL-405 and a 650 nm ps diode laser of Becker & Hickl. The pulse width of the lasers was 70 to 80 ps, i.e. much shorter that the detector response. The measurements show that the TTS is not only much wider than specified, there is also a considerable change with the wavelength, and, still worse, with the count rate. Therefore the SPCM-AQR cannot be used for fluorescence lifetime measurements.

Interestingly, an older SPCM-AQR had a smaller count-rate dependence. Fig. 40 shows the TCSPC response of an SPCM-AQR-14 (dark count class < 40 cps) manufactured in 1999. Although the shift with the count rate is still too large for fluorescence lifetime experiments, it is much smaller than for the new device.


Fig. 41: SPCM-AQR-12, histogram of times between photons

 

The afterpulse probability of the SPCM-AQR is low enough for correlation experiments down to a few 100 ns, fig. 41. An inconvenience of the non-fibre version of the SPCM-AQR is that it is almost impossible to attach it to an optical system without getting daylight into the optical path. A standard optical adapter, e.g. a C-mount thread around the photodiode, would simplify the optical setup considerably.

The conclusion is that the SPCM-AQR is an excellent detector for fluorescence correlation spectroscopy and high efficiency steady state photon counting but not applicable to fluorescence lifetime measurements. This is disappointing, particularly because state-of-the-art TCSPC techniques allow for simultaneous FCS / lifetime measurements which are exceptionally useful to investigate conformational changes in protein-dye complexes, single-molecule FRET and diffusion processes in living cells. Currently the only solution for these applications is to use PMT detectors, i.e. the R3809U MCP, the H7422 or the R5900 which, of course, means to sacrifice some efficiency.

spcmicr1.pdf
Full Text
 

 

 

 

 
© 1993-2010
Becker & Hickl GmbH    
Nahmitzer Damm 30
12277 Berlin
Geschäftsführer: Dr. Wolfgang Becker, Helmut Hickl
Amtsgericht Berlin-Charlottenburg HRB 50729
Steuernummer: 29 410 05494
USt-IDNr: 164939770
   email info@becker-hickl.com
   fon +49 (30) 787 56 32
   fax +49 (30) 787 57 34